Devices to be operated: Needs access to the nurse call system and also wants to control a television.
The patient did not have the strength to operate a standard 'light pressure' switch but could reliably move her hand. Thus an Electronic Touch Switch was used with an extension sensor mounted on a boom arm stand beside the bed. This switch is so sensitive that any contact with bare skin will activate it - no pressure is required.
Did you know that you can subscribe to our email
list
and receive updates when we add new articles to the TS Files. Simply click
the following link and type 'TS Files' as the subject. Feel free to suggest
topics for future TS Files in the email body and make sure to provide your
email address if you are not sending this from your own computer. Click
to email and subscribe
(You can unsubscribe at any time)
A Nurse Call Timer was adapted to interface with the existing nurse call system - so that it could be plugged into the wall as a replacement for the standard nurse call pendant.

The Electronic Touch Switch plugs into the Nurse Call Timer (Call Interface) which in turn plugs into the call system.
In order to give the patient control over a television, the ECU output from the Nurse Call Timer is connected to a scanning Television Controller.
Operation:
Brief activation's of the switch will activate the scanning Television Controller,
while a longer sustained activation will trigger the nurse call alarm.